Safety of Screening Procedures With Hand-Held Metal Detectors Among Patients With Implanted Cardiac Rhythm Devices A Cross-sectional Analysis

Jilek, Clemens and Tzeis, Stylianos and Vrazic, Hrvoje and Semmler, Verena and Andrikopoulos, Georgios and Reents, Tilko and Fichtner, Stephanie and Ammar, Sonia and Rassias, Ioannis and Theodorakis, Georgios and Weber, Stefan and Hessling, Gabriele and Deisenhofer, Isabel and Kolb, Christof (2011) Safety of Screening Procedures With Hand-Held Metal Detectors Among Patients With Implanted Cardiac Rhythm Devices A Cross-sectional Analysis. ANNALS OF INTERNAL MEDICINE, 155 (9). 587-U53. ISSN 0003-4819, 1539-3704

Full text not available from this repository. (Request a copy)

Abstract

Background: Case reports suggest that the hand-held metal detectors used for security screening generate electromagnetic fields that may interfere with pacemaker and implantable cardioverter-defibrillator (ICD) function. Objective: To assess changes in function of pacemakers and ICDs after exposure to hand-held metal detectors. Design: Cross-sectional study. Setting: Two medical centers in Europe. Patients: 388 patients (209 with pacemakers and 179 with ICDs) presenting for routine follow-up of device function between September 2009 and December 2010. Measurements: Abnormalities on electrocardiography suggestive of rhythm device malfunction (pacing inhibition, loss of capture, inappropriate mode switch, ventricular oversensing, and spontaneous reprogramming) after 30 seconds of exposure to 2 widely used hand-held metal detectors with a maximal electromagnetic flux density of 6.3 mu T. Results: No change in device function, including pacing or sensing abnormalities or device reprogramming, was observed in any patient. Limitations: The study included a convenience sample of patients, and the number of different device models tested was small. Testing was conducted in 2 clinic settings. Conclusion: Hand-held metal detectors did not affect the function of pacemakers or ICDs in this sample. The use of hand-held metal detectors for security screening is probably safe for patients with pacemakers and ICDs, but these findings require confirmation. Primary Funding Source: None.

Item Type: Article
Uncontrolled Keywords: CARDIOVERTER-DEFIBRILLATORS; ELECTROMAGNETIC-INTERFERENCE; PERMANENT PACEMAKERS; MEDIA PLAYERS; TELEMETRY;
Subjects: 600 Technology > 610 Medical sciences Medicine
Divisions: Medicine > Lehrstuhl für Innere Medizin II
Depositing User: Dr. Gernot Deinzer
Date Deposited: 27 May 2020 09:59
Last Modified: 27 May 2020 09:59
URI: https://pred.uni-regensburg.de/id/eprint/19914

Actions (login required)

View Item View Item