Dobler, Barbara and Groeger, Christian and Treutwein, Marius and Alvarez-Moret, Judith and Goetzfried, Thomas and Weidner, Karin and Haertl, Petra and Koelbl, Oliver (2011) Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment. RADIOTHERAPY AND ONCOLOGY, 99 (1). pp. 86-89. ISSN 0167-8140, 1879-0887
Full text not available from this repository. (Request a copy)Abstract
Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan (R) v3.3 with Elekta's Mosaiq (R) v1.6 and SynergyS (R) linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 99 (2011) 86-89