Berto, Raphaela and Strutz, Juergen (2017) Destruction of microsurgical devices by sterilisation. LARYNGO-RHINO-OTOLOGIE, 96 (11). pp. 774-779. ISSN 0935-8943, 1438-8685
Full text not available from this repository.Abstract
Hospital facilities issue numerous risk announcements on corrosion, deformation or premature wearout of medical devices every year. As there is yet little data on the impact of reprocessing on the quality and durability of microsurgical instruments, this paper aims at evaluating the effects of the reprocessing on microsurgical instruments. Material and Methods 22 brand new microsurgical instruments for stapes surgery were being reprocessed 30 times without being used for surgery or other purposes in the interim time. After each reprocessing the instruments were examined macroscopicly and microscopicly. The results were portrayed in a photo documentation and analysed on that basis. Results Almost all devices showed mechanical damage caused by the reprocessing procedure. The increasing deterioration was often associated with missing protective caps. Furthermore contaminations and stains were apparent in several cases. Conclusions The findings illustrate that careful handling of delicate surgical devices during reprocessing is vital. They also highlight problems of protective caps. As an alternative going forward it should be considered to store microsurgical instruments statically in special racks.
| Item Type: | Article |
|---|---|
| Uncontrolled Keywords: | ; reprocessing; damage; microsurgical devices; protective caps; stapes surgery; photo documentation |
| Subjects: | 600 Technology > 610 Medical sciences Medicine |
| Divisions: | Medicine > Lehrstuhl für Hals-Nasen-Ohren-Heilkunde |
| Depositing User: | Dr. Gernot Deinzer |
| Date Deposited: | 14 Dec 2018 13:19 |
| Last Modified: | 26 Feb 2019 08:51 |
| URI: | https://pred.uni-regensburg.de/id/eprint/1915 |
Actions (login required)
![]() |
View Item |