Weber, Stefan and Jeron, Andreas and Schneider, Hans J. and Muders, Frank (2006) Efficacy of patient activated antitachycardia pacing therapy using the Medtronic AT500 pacemaker. EUROPACE, 8 (6). pp. 413-415. ISSN 1099-5129,
Full text not available from this repository. (Request a copy)Abstract
A 28-year-old female with corrected anomalous pulmonary venous drainage presented with multiple atrial arrhythmias many years later. Ablation techniques eliminated most of the arrhythmias except atrial tachycardia with 1:1 AV conduction. A dual-chamber pacemaker with antiatrial tachycardia features was implanted and was shown to be effective in arrhythmia control when the standard algorithm was overridden by an external patient activation device.