Lenhart, Markus and Voelk, Markus and Manke, Christoph and Nitz, Wolfgang R. and Strotzer, Michael and Feuerbach, Stefan and Link, Johann (2000) Stent appearance at contrast-enhanced MR Angiography: In vitro examination with 14 stents. RADIOLOGY, 217 (1). pp. 173-178. ISSN 0033-8419
Full text not available from this repository. (Request a copy)Abstract
PURPOSE: To evaluate signal intensity changes influencing assessment of stent patency at contrast material-enhanced magnetic resonance (MR) angiography. MATERIALS AND METHODS: By using an in vitro model, 14 stents-nine nitinol, one tantalum, two stainless steel, and two cobalt alloy-were investigated regarding their appearance at MR imaging. A vascular phantom consisting of tubes filled with 2.00 mmol/L gadopentetate dimeglumine in saline solution was studied in different orientations within the magnetic field. Imaging was performed with a fast three- dimensional gradient-echo sequence (4.70/1.89 [repetition time msec/echo time msec]). Relative signal intensity reduction within the stents and the degree of artificial narrowing of the stent lumen were calculated. RESULTS: The stent lumen was visible within 13 stents. A total signal void inside the stent lumen appeared in only one cobalt alloy stent. Artificial narrowing of the diameter was less than 33% in 10 of 14 stents. The tantalum stent and four nitinol stents seemed best suited for contrast-enhanced MR angiography. A bandlike artifact occurred at the ends of the stents when positioned along the readout direction. CONCLUSION: To differentiate between artifacts and stenoses, knowledge of the degree of signal intensity reduction and artificial lumen narrowing within vascular stents is essential. Stent geometry, relative orientation to the magnetic field, and alloy composition influence signal intensity alteration within the stent lumen.
| Item Type: | Article |
|---|---|
| Uncontrolled Keywords: | ARTIFACTS; PLACEMENT; VISUALIZATION; INTERVENTIONS; CATHETER; ANEURYSM; arteries, grafts and prostheses; arteries, MR; magnetic resonance(MR), artifact; magnetic resonance(MR), vascular studies; stents and prostheses |
| Subjects: | 600 Technology > 610 Medical sciences Medicine |
| Divisions: | Medicine > Lehrstuhl für Röntgendiagnostik |
| Depositing User: | Dr. Gernot Deinzer |
| Date Deposited: | 23 Mar 2022 07:06 |
| Last Modified: | 23 Mar 2022 07:06 |
| URI: | https://pred.uni-regensburg.de/id/eprint/42193 |
Actions (login required)
![]() |
View Item |