Hohenleutner, Ulrich and Wenig, M. and Walther, T. and Baeumler, W. and Landthaler, Michael (1998) Treatment of ectatic leg veins with a flashlamp-pumped pulsed dye laser with 1,5 ms pulse length. HAUTARZT, 49 (7). pp. 560-565. ISSN 0017-8470,
Full text not available from this repository. (Request a copy)Abstract
A new flashlamp-pumped pulsed dye laser with 1,5 ms pulse length and tunable wavelengths between 585 and 600 nm (Candela ScleroPLUS) is now available for the treatment of ectatic leg veins.We evaluated 232 single test treatments using 595 and 600 nm wavelength in 75 test persons. For smaller vessels up to a diameter of 0,5 mm, treatments at 16 and 18 J/cm(2) showed good results in 60 and 82% respectively;for Vessels up to 1 mm in 27 and 33%.Treatment at 595 nm was a little more effective than at 600 nm. Larger vessels could not be treated effectively.The main side effects consisted in hyper- (34%) or hypopigmentation (30%).Thus, in addition to the usual sclerotherapy, this laser is a useful device for the treatment of leg vein ectasia.
| Item Type: | Article |
|---|---|
| Uncontrolled Keywords: | PORT-WINE STAINS; LOWER-EXTREMITY; THERAPY; TELANGIECTASIAS; VARICOSITIES; OPTICS; ARGON; SKIN; ectatic leg veins; sclerotherapy; flashlamp-pumped pulsed dye laser; selective photothermolysis |
| Subjects: | 600 Technology > 610 Medical sciences Medicine |
| Divisions: | Medicine > Lehrstuhl für Dermatologie und Venerologie |
| Depositing User: | Dr. Gernot Deinzer |
| Date Deposited: | 09 May 2023 07:09 |
| Last Modified: | 09 May 2023 07:09 |
| URI: | https://pred.uni-regensburg.de/id/eprint/49695 |
Actions (login required)
![]() |
View Item |