Accuracy of quantitative digital subtraction radiography for determining changes in calcium mass in mandibular bone: an in vitro study

Christgau, Michael and Hiller, Karl-Anton and Schmalz, Gottfried and Kolbeck, C. and Wenzel, A. (1998) Accuracy of quantitative digital subtraction radiography for determining changes in calcium mass in mandibular bone: an in vitro study. JOURNAL OF PERIODONTAL RESEARCH, 33 (3). pp. 138-149. ISSN 0022-3484, 1600-0765

Full text not available from this repository. (Request a copy)

Abstract

The aim of this in vitro study was to determine the accuracy of digital subtraction radiography (DSR) to detect small changes in calcium mass in alveolar bone adjacent to tooth roots. In each of 4 dried porcine mandible segments, one interproximal and one buccal "defect" region was defined adjacent to a premolar root. A series of cortical and cancellous bone slices with a 50 mu m-stepwise increasing thickness (0-5000 mu m) mere attached to the mandible segments covering the respective "defect" region. Standardized radiographs were quantitatively assessed for density changes using DSR. After dissolving each bone slice in hydrochloric acid, its calcium concentration was photometrically determined. For each bone slice, the mean calcium mass covering a single pixel of the subtraction image was calculated. The Wilcoxon signed-rank test and the Mann-Whitney U-test were used for statistical analysis (alpha=0.05). A strong linear correlation (r(2)=0.86-1.00; p less than or equal to 0.001) was found between the thickness of the bone slices and their calcium mass. Cortical bone showed a 3.5 times higher mean calcium mass/pixel than cancellous bone. Furthermore, a strong linear correlation (r(2)=0.63-1.00; p less than or equal to 0.001) was found between the mean calcium mass per image pixel and the radiographic density changes. Neither the bone type nor the "defect" localization had a significant influence on radiographic density changes caused by changes in calcium mass. A change in mean calcium mass per image pixel of 0.1-0.15 mg was necessary to be detected by DSR. In conclusion, this study revealed a high accuracy of DSR to detect small changes in calcium mass in alveolar cortical and cancellous bone.

Item Type: Article
Uncontrolled Keywords: ALVEOLAR BONE; STANDARDIZED RADIOGRAPHS; SIMULATED OSTEOPOROSIS; I-125 ABSORPTIOMETRY; PROJECTION GEOMETRY; IN-VITRO; LESIONS; ERRORS; REPRODUCIBILITY; subtraction radiography; quantitative analysis; calcium; alveolar bone
Subjects: 600 Technology > 610 Medical sciences Medicine
Divisions: Medicine > Lehrstuhl für Zahnerhaltung und Parodontologie
Depositing User: Dr. Gernot Deinzer
Date Deposited: 12 Sep 2023 07:14
Last Modified: 12 Sep 2023 07:14
URI: https://pred.uni-regensburg.de/id/eprint/49980

Actions (login required)

View Item View Item