LOW BLANK RHENIUM ISOTOPE RATIO DETERMINATIONS BY V2O5 COATED NICKEL FILAMENTS USING NEGATIVE THERMAL IONIZATION MASS-SPECTROMETRY

WALCZYK, T and HEBEDA, EH and HEUMANN, KG (1994) LOW BLANK RHENIUM ISOTOPE RATIO DETERMINATIONS BY V2O5 COATED NICKEL FILAMENTS USING NEGATIVE THERMAL IONIZATION MASS-SPECTROMETRY. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 130 (3). pp. 237-246. ISSN 0168-1176,

Full text not available from this repository.

Abstract

Rhenium isotope ratio determinations are, in principle, possible by negative thermal ionization mass spectrometry (NTI-MS). Relatively high rhenium blanks from the commonly-used filament materials prevent accurate isotope ratio determinations, especially for small rhenium sample amounts which are of importance, for example, in geochronology in connection with the Re/Os dating method. Platinum and nickel filaments were tested by different preparation techniques to reduce the rhenium blank contribution from the filament material. The lowest rhenium blank of less than 1 pg was achieved by coating nickel filaments with V2O5 prior to degassing under high vacuum conditions at 850 degrees C. Obviously, the vanadium-nickel oxide layer formed on the surface of the filament during this process prevents further emission of rhenium ions from the filament material. Using Ba(OH)(2) for the enhancement of negative thermal ions, 1 ng of rhenium resulted in ion currents at the detector side of about 10(-11) A with an ionization efficiency of up to 20%. The Re-185/Re-187 isotope ratio of a sample of natural isotopic composition could be determined to be 0.59818 +/- 0.00026 with a relative precision of 0.04%. The isotope ratio determination for an Re-187 spike was comparable in precision but the relative standard deviation of an Re-185 spike was significantly higher, which could be explained by mass fractionations of oxygen in the measured ReO4- ion. The ReO4- ion is about 200 to 2500 times more abundant than the only other detectable rhenium ion in NTI-MSReO(3)(-). The ReO4-/ReO3- ratio decreases with increasing temperature. By the low blank NTI-MS technique described in this work, more precise and accurate determinations of the rhenium isotope ratio and the rhenium concentration by isotope dilution analysis from nanogramme samples are possible.

Item Type: Article
Uncontrolled Keywords: OSMIUM; ELEMENTS; RHENIUM; ISOTOPE RATIO; NEGATIVE THERMAL IONIZATION MASS SPECTROMETRY; NICKEL FILAMENTS
Depositing User: Dr. Gernot Deinzer
Last Modified: 19 Oct 2022 08:40
URI: https://pred.uni-regensburg.de/id/eprint/53465

Actions (login required)

View Item View Item